KNIFE
Report
- Report Number
- 2523835-2013-00049
- Event Type
- Malfunction
- Date Received
- April 15, 2013
- Report Date
- April 4, 2013
- Manufacturer
- ALCON PRECISION DEVICE - SINKING SPRING
- Product Code
- HNN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
NO SAMPLE WERE RETURNED FOR EVALUATION; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MFR'S ACCEPTANCE CRITERIA. BECAUSE A SAMPLE WAS NOT RETUNED AND NO EVIDENCE OF NON-CONFORMITY COULD BE FOUND IN THE LOT RECORD REVIEW, THE ROOT CAUSE FOR THE DULL KNIFE EXPERIENCED BY THE CUSTOMER CAN NOT BE DETERMINED. SOME POTENTIAL CAUSES ARE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SETUP. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MFG. ANY DEFECTS SUCH AS DAMAGED TIPS AND CUTTING EDGES, ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).
AN OPHTHALMIC SURGEON REPORTED THE BLADE OF A KNIFE WAS DULL DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 161818 | KNIFE | MANUAL OPHTHALMIC SURGICAL INSTRUMENT | HNN | ALCON PRECISION DEVICE - SINKING SPRING | 8065982465 | 889627M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |