FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 3063732 · Received March 28, 2013

Report

Report Number
1824206-2013-01782
Event Type
Malfunction
Date Received
March 28, 2013
Date of Event
March 1, 2013
Report Date
March 1, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE POWER CORD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECH ALLEGED THE BED DID NOT PASS THE GROUND RESISTANCE TEST. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
127406 TOTALCARE BARIATRIC BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1