FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 3063689 · Received April 12, 2013

Report

Report Number
2242352-2013-00352
Event Type
Malfunction
Date Received
April 12, 2013
Date of Event
March 18, 2013
Report Date
March 26, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, SEAL, AND ANCHOR TAB WERE LOCATED INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK WAS LOCKED AND THE WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL FAILED TO LOAD PROPERLY AS THE SEAL DID NOT FOLD PROPERLY AND REMAINED INSIDE OF THE LOADING DEVICE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
159617 HS III PROXIMAL SEAL SYSTEM 3.8MM CLAMPLESS BEATING HEART GEI MAQUET CARDIOVASCULAR, LLC HSK-3038 25068306

Patients

Seq Age Sex Outcome Treatment
1 NA