CONSTELLATION VISION SYSTEM

MDR report key: 3063670 · Received April 12, 2013

Report

Report Number
2028159-2013-00684
Event Type
Malfunction
Date Received
April 12, 2013
Date of Event
March 22, 2013
Report Date
March 22, 2013
Manufacturer
ALCON ¿ IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K101285
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EXAMINED THE SYSTEM AND CONFIRMED THE PROBLEM REPORTED. THE AIR HOSE ASSEMBLY WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVAL AND NO ADDITIONAL INFO PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE WAS A NON CONFORMING AIR HOSE ASSEMBLY. (B)(4).

Additional Manufacturer Narrative · 1

THE COMPANY REP EXAMINED THE SYSTEM AND CONFIRMED THE PROBLEM REPORTED. THE AIR HOSE ASSEMBLY WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVAL AND NO ADDITIONAL INFO PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE WAS A NON CONFORMING AIR HOSE ASSEMBLY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT "DECREASING GAS PRESSURE" DISPLAYED DURING A VITRECTOMY PROCEDURE. THE CUSTOMER CONFIRMED THAT AIR WAS LEAKING FROM THE GAS HOSE. THE STAFF HELD THE GAS HOSE TO COMPLETE THE PROCEDURE. THERE WAS NO HARM TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157405 CONSTELLATION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON ¿ IRVINE TECHNOLOGY CENTER LXT-JAPAN NA

Patients

Seq Age Sex Outcome Treatment
1