FDA Adverse Event Malfunction Summary report: N

SETROX S 53

MDR report key: 3063666 · Received April 17, 2013

Report

Report Number
1028232-2013-00991
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
December 21, 2012
Report Date
April 8, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE OUTER AND INNER COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. COAGULATED BLOOD WAS FOUND IN THE LUMEN OF THE LEAD. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD WAS EXPLANTED AND REPLACED DUE TO LOSS OF CAPTURE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166173 SETROX S 53 PACER LEAD NVN BIOTRONIK SE & CO. KG 350974

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization