PUMP IN STYLE (ITEM # UNK)
Report
- Report Number
- 1419937-2013-00208
- Event Type
- Malfunction
- Date Received
- April 12, 2013
- Date of Event
- March 14, 2013
- Report Date
- March 21, 2013
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
ATTEMPTS TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL. THE PRODUCT INVOLVED IN THE COMPLAINT HAS NOT BEEN RECEIVED FOR TESTING/ANALYSIS. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. HOWEVER, THE CUSTOMER REPORTED THE HOUSING CAME APART AND THE INNER ELECTRONICS WERE EXPOSED. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.
THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE POWER SUPPLY FOR HER 10 MONTH OLD BREAST PUMP HAS CRACKED AND FALLEN APART EXPOSING INNER ELECTRONICS. NO INJURIES WERE SUSTAINED AS A RESULT OF THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159329 | PUMP IN STYLE (ITEM # UNK) | HGX | MEDELA, INC. | 57XXX/9207010 | REV M - 1212 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |