FDA Adverse Event Malfunction Summary report: N

REMB SAG SAW

MDR report key: 3063529 · Received April 17, 2013

Report

Report Number
0001811755-2013-00821
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 27, 2013
Report Date
March 27, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE MOTOR AND THE ROTOR. ADDITIONALLY THE BOOT ON THE SAGHEAD WAS WORN.

Description of Event or Problem · 1

THE REMB SAG SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
164762 REMB SAG SAW DRILLS, BURRS, TREPHINES & ACC HBE STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1