FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM

MDR report key: 3063395 · Received April 17, 2013

Report

Report Number
1226181-2013-00178
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 19, 2013
Report Date
March 19, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC-BROOKFIELD
Product Code
CGA
PMA / PMN Number
K051087
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED GLUCOSE RESULT IS SAMPLE INTEGRITY CAUSED BY AN ALIQUOT PROBE DRAIN OVERFLOW. THE SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO PERFORM MAINTENANCE ACTIVITIES TO CLEAN THE ALIQUOT PROBE DRAIN. SUBSEQUENT QC WAS WITHIN RANGE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATION. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

FALSELY ELEVATED GLUCOSE RESULTS WERE OBTAINED ON A PATIENT SAMPLE. THE PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE TEST WAS RE-RUN ON AN ALTERNATE DIMENSION VISTA INSTRUMENT AND A LOWER RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED GLUCOSE RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
165758 DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM DIMENSION® VISTA® CLINICAL CHEMISTRY SYSTEM CGA SIEMENS HEALTHCARE DIAGNOSTICS INC-BROOKFIELD DIMENSION VISTA 1500

Patients

Seq Age Sex Outcome Treatment
1