FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2

MDR report key: 3063342 · Received April 11, 2013

Report

Report Number
2242352-2013-00342
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 21, 2013
Report Date
March 25, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE AND NO EVIDENCE OF BLOOD ON THE DEVICE. VISUAL INSPECTION DETERMINED THAT THE HEATER WIRE WAS BENT AND DETACHED FROM THE TIP OF THE HOT JAW. BASED UPON THE REVIEW CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE METAL TIP ON THE HEMOPRO 2 JAW WAS BENT. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154753 VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25071718

Patients

Seq Age Sex Outcome Treatment
1 NA