HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-09526
- Event Type
- Malfunction
- Date Received
- April 17, 2013
- Date of Event
- February 26, 2013
- Report Date
- April 2, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE AN UNEXPECTED HIGH ULTRA-FILTRATION (UF) FOR THIS THERAPY COMPARED TO OTHER THERAPIES. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 22:42:42. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 1723ML, INDICATING THE HOME PATIENT (HP) DRAINED 1723ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1800ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165382 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |