FDA Adverse Event Malfunction Summary report: N

OPTIFLUX 160NRE DIALYZER FINISHED ASSY.

MDR report key: 3063297 · Received March 1, 2013

Report

Report Number
1713747-2013-00049
Event Type
Malfunction
Date Received
March 1, 2013
Date of Event
January 17, 2013
Report Date
February 4, 2013
Manufacturer
OGDEN MANUFACTURING
Product Code
FJI
PMA / PMN Number
K002761
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

A HEMODIALYSIS USER FACILITY HAS REPORTED THAT DURING TREATMENT, A BLOOD LEAK OCCURRED. THE LEAK WAS VISUALLY OBSERVED AND TEST STRIPS WERE USED TO CONFIRM THE BLOOD LEAK. ESTIMATED BLOOD LOSS WAS 200CC'S. PATIENT HAD NO ILL EFFECTS. SAMPLE HAS NOT BEEN RETURNED TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89890 OPTIFLUX 160NRE DIALYZER FINISHED ASSY. HEMODIALYSIS DIALYZER FJI OGDEN MANUFACTURING 12PU03007

Patients

Seq Age Sex Outcome Treatment
1 FRESENIUS 2008K MACHINE