FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, SINGLE CONN./EXT. DL

MDR report key: 3063276 · Received March 1, 2013

Report

Report Number
8030665-2013-00121
Event Type
Malfunction
Date Received
March 1, 2013
Date of Event
February 4, 2013
Report Date
February 5, 2013
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE COMPLAINT IS CONFIRMED WITH 6 PINHOLES ON THE CASSETTE FILM. AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS PATIENT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER DURING TREATMENT. THE PATIENT'S NURSE REMOVED THE CASSETTE AND NOTICED FLUID INSIDE THE CYCLER DOOR. PATIENT HAD NO ILL EFFECTS. SAMPLE IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89888 LIBERTY CYCLER SET, SINGLE CONN./EXT. DL LIBERTY DIALYSIS CYCLER TUBING FKX REYNOSA MANUFACTURING 12LR08035

Patients

Seq Age Sex Outcome Treatment
1 LIBERTY DIALYSIS CYCLER