FDA Adverse Event Malfunction Summary report: N

20 DEGREE ANGLE DRIVER

MDR report key: 3063208 · Received April 11, 2013

Report

Report Number
1045834-2013-01591
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 15, 2013
Report Date
March 18, 2013
Manufacturer
THE ANSPACH EFFORT, INC.
Product Code
HBE
PMA / PMN Number
K042783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED BY ANSPACH. THE DEVICE WAS EVALUATED AND FAILED THE TEMPERATURE ACCEPTANCE TEST (OVER HIGH LIMIT). EVIDENCE SUGGESTS THIS WAS DUE TO USAGE/WEAR OVER TIME. THE EVENT WAS CONFIRMED. IF ADD'L INFO IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM THE USA STATING THAT THE DEVICE WAS "OVERHEATING" DURING TESTING. THE DEVICE WAS NOT USED IN SURGERY. THERE WERE NO INJURIES OR MEDICAL INTERVENTION REPORTED. THERE WAS NO ADD'L INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154174 20 DEGREE ANGLE DRIVER HBE THE ANSPACH EFFORT, INC.

Patients

Seq Age Sex Outcome Treatment
1