FDA Adverse Event Malfunction Summary report: N

SONOSURG-G2

MDR report key: 3063189 · Received February 27, 2013

Report

Report Number
8010047-2013-00046
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 12, 2013
Report Date
December 25, 2015
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVAL. THE EVAL CONFIRMED THAT THE FUSES OF THE SUBJECT DEVICE BLEW AND FET (FIELD-EFFECT TRANSISTOR) OF AMPLIFIER SUBSTRATE WAS DAMAGED. THE DAMAGE OF FET OCCURRED. BUT IT WAS NOT KNOWN EXACTLY CAUSE. THE MFG HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED UPON THE EVAL RESULT OF THE SUBJECT DEVICE BY OMSC, SINCE THE INCIDENTAL MALFUNCTION OF FET OCCURRED OR THERE WAS SOMETHING THE WRONG WITH THE ELECTRICAL POWER ENVIRONMENT OF FACILITY, FET WAS DAMAGED AND THE FUSES BLEW. THE SONOSURG-G2 INSTRUCTION MANUAL ALREADY STATES. WARNING: PREPARE A SPARE INSTRUMENT AS A BACKUP TO ENSURE THAT THE PROCEDURE CAN BE COMPLETED W/O COMPLICATIONS IN CASE OF A MALFUNCTION. WARNING: IF YOU SUSPECT ANY ABNORMALITY WHILE USING THIS PRODUCT, IMMEDIATELY STOP THE PROCEDURE. REFER TO CHAPTER 7, "TROUBLESHOOTING" AND THE INSTRUCTION MANUAL OF THE SONOSURG HAND PIECE BEING USED. IF THE INFO GIVEN THERE DOES NOT ELIMINATE THE ABNORMALITY, TERMINATE THE USE OF THE EQUIPMENT AND CONTACT OLYMPUS. IF ADD'L INFO IS REC'D, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) PERFORMED A MDR RETROSPECTIVE REVIEW AND OMSC FOUND THAT THIS SUPPLEMENTAL REPORT IS REQUIRED ON DECEMBER 3, 2015. THIS IS A SUPPLEMENTAL REPORT FOR MFR REPORT #8010047-2013-00046 TO PROVIDE DEVICE EVALUATION RESULTS. OLYMPUS PERFORMED FURTHER INVESTIGATION FOR THE CAUSE OF THE FET (FIELD EFFECT TRANSISTOR) DAMAGE AND THE DAMAGED FET HAD EXCESSIVE CURRENT LEAKAGE. THE AMOUNT OF THE LEAKAGE CURRENT DEPENDS ON THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET. OLYMPUS IMPLEMENTED A COUNTERMEASURE FOR THIS PHENOMENON TO REDUCE THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET.

Description of Event or Problem · 1

WHEN A PHYSICIAN ATTEMPTED TO USE THE SUBJECT DEVICE FOR A LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY, THE FOOTSWITCH FOR OUTPUT DIDN'T WORK. THE PHYSICIAN REPLACED THE SUBJECT DEVICE WITH A SIMILAR DEVICE. THE PROCEDURE WAS COMPLETED AS SCHEDULED. THERE WAS NO PT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85291 SONOSURG-G2 ULTRASONIC SURGICAL SYSTEM LFL OLYMPUS MEDICAL SYSTEMS CORP. SONOSURG-G2 NA

Patients

Seq Age Sex Outcome Treatment
1