SONOSURG-G2
Report
- Report Number
- 8010047-2013-00046
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- February 12, 2013
- Report Date
- December 25, 2015
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVAL. THE EVAL CONFIRMED THAT THE FUSES OF THE SUBJECT DEVICE BLEW AND FET (FIELD-EFFECT TRANSISTOR) OF AMPLIFIER SUBSTRATE WAS DAMAGED. THE DAMAGE OF FET OCCURRED. BUT IT WAS NOT KNOWN EXACTLY CAUSE. THE MFG HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED UPON THE EVAL RESULT OF THE SUBJECT DEVICE BY OMSC, SINCE THE INCIDENTAL MALFUNCTION OF FET OCCURRED OR THERE WAS SOMETHING THE WRONG WITH THE ELECTRICAL POWER ENVIRONMENT OF FACILITY, FET WAS DAMAGED AND THE FUSES BLEW. THE SONOSURG-G2 INSTRUCTION MANUAL ALREADY STATES. WARNING: PREPARE A SPARE INSTRUMENT AS A BACKUP TO ENSURE THAT THE PROCEDURE CAN BE COMPLETED W/O COMPLICATIONS IN CASE OF A MALFUNCTION. WARNING: IF YOU SUSPECT ANY ABNORMALITY WHILE USING THIS PRODUCT, IMMEDIATELY STOP THE PROCEDURE. REFER TO CHAPTER 7, "TROUBLESHOOTING" AND THE INSTRUCTION MANUAL OF THE SONOSURG HAND PIECE BEING USED. IF THE INFO GIVEN THERE DOES NOT ELIMINATE THE ABNORMALITY, TERMINATE THE USE OF THE EQUIPMENT AND CONTACT OLYMPUS. IF ADD'L INFO IS REC'D, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) PERFORMED A MDR RETROSPECTIVE REVIEW AND OMSC FOUND THAT THIS SUPPLEMENTAL REPORT IS REQUIRED ON DECEMBER 3, 2015. THIS IS A SUPPLEMENTAL REPORT FOR MFR REPORT #8010047-2013-00046 TO PROVIDE DEVICE EVALUATION RESULTS. OLYMPUS PERFORMED FURTHER INVESTIGATION FOR THE CAUSE OF THE FET (FIELD EFFECT TRANSISTOR) DAMAGE AND THE DAMAGED FET HAD EXCESSIVE CURRENT LEAKAGE. THE AMOUNT OF THE LEAKAGE CURRENT DEPENDS ON THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET. OLYMPUS IMPLEMENTED A COUNTERMEASURE FOR THIS PHENOMENON TO REDUCE THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET.
WHEN A PHYSICIAN ATTEMPTED TO USE THE SUBJECT DEVICE FOR A LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY, THE FOOTSWITCH FOR OUTPUT DIDN'T WORK. THE PHYSICIAN REPLACED THE SUBJECT DEVICE WITH A SIMILAR DEVICE. THE PROCEDURE WAS COMPLETED AS SCHEDULED. THERE WAS NO PT HARM REPORTED.