MEDFUSION 3500 SYRINGE INFUSION PUMP

MDR report key: 3063076 · Received February 18, 2013

Report

Report Number
2183502-2013-00036
Event Type
Malfunction
Date Received
February 18, 2013
Date of Event
January 18, 2013
Report Date
February 15, 2013
Manufacturer
SMITHS MEDICAL MD
Product Code
FRN
PMA / PMN Number
K040899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. TESTING INDICATED THAT THE POSITION POTENTIOMETER WAS NON-FUNCTIONAL DUE TO A BROKEN TAB. THE POSITION POTENTIOMETER WAS REPLACED. FOLLOWING REPAIR, THE DEVICE PASSED ALL FUNCTION AND DELIVERY TESTING.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH". NO PT INJURY OR ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72213 MEDFUSION 3500 SYRINGE INFUSION PUMP SYRINGE INFUSION PUMP FRN SMITHS MEDICAL MD 3500 M83761