FDA Adverse Event Injury Summary report: N

THERAPY COOL FLEX, 1304-CF-7-0.5(5) 2-M-TE4BE1EB

MDR report key: 3062927 · Received April 12, 2013

Report

Report Number
2030404-2013-00029
Event Type
Injury
Date Received
April 12, 2013
Date of Event
March 20, 2013
Report Date
March 20, 2013
Manufacturer
ST. JUDE MEDICAL, INC. (AF-IRVINE)
Product Code
OAD
PMA / PMN Number
P060019
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE DEVICE HISTORY RECORD WAS NOT POSSIBLE SINCE THE LOT NUMBER WAS UNAVAILABLE. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED. PER THE IFU, VASCULAR PERFORATION IS A KNOWN INHERENT RISK OF ANY ELECTRODE PLACEMENT.

Description of Event or Problem · 1

RELATED TO MFR REFERENCE 3005188751-2013-00054, 3005188751-2013-00056. DURING A VENTRICULAR TACHYCARDIA ABLATION PROCEDURE USING A THERAPY COOL FLEX ABLATION CATHETER, A CARDIAC PERFORATION OCCURRED. DURING SETUP AND WHILE CREATING GEOMETRY WITH THE VELOCITY SYSTEM, THE PHYSICIAN ENCOUNTERED SEVERAL ERROR MESSAGES AND A CATHETER DRIFT. RESPIRATORY COMPENSATION DID NOT RESOLVE THE CATHETER DRIFT, SO THE PHYSICIAN CONTINUED THE PROCEDURE USING FLUOROSCOPY AS WELL. THE PATIENT BECAME HYPOTENSIVE AND THEIR RESPIRATORY PATTERN CHANGED. ULTRASOUND CONFIRMED A PERICARDIAL EFFUSION. THE PHYSICIAN STOPPED THE PROCEDURE AND A PERICARDIOCENTESIS WAS PERFORMED WHICH STABILIZED THE PATIENT. NO FURTHER INTERVENTION WAS REQUIRED. THE PHYSICIAN BELIEVES THE CARDIAC PERFORATION OCCURRED DURING TRANSSEPTAL PUNCTURE OR WHILE MANIPULATING THE CATHETER NEAR THE MITRAL VALVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157834 THERAPY COOL FLEX, 1304-CF-7-0.5(5) 2-M-TE4BE1EB CARDIAC ABLATION PERCUTANEOUS CATHETER OAD ST. JUDE MEDICAL, INC. (AF-IRVINE) A088015 UNK

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention BARD DYNAMIC DECA CATHETER (201101/LOT UNKNOWN)| BRK TRANSSEPTAL NEEDLE (407201/LOT UNKNOWN)| SUPREME EP CATHETER (402010/LOT UNKNOWN)| SUPREME EP CATHETER (401442/LOT UNKNOWN)| SWARTZ BRAIDED INTRODUCER (407453/LOT UNKNOWN)| ENSITE VELOCITY SYSTEM (V3.0.1.1)