ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-04398
- Event Type
- Malfunction
- Date Received
- April 17, 2013
- Report Date
- March 22, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 05/31/2013 DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/01/2013 WITH THE FOLLOWING FINDINGS:INVESTIGATION REVEALED A DIM/FADED AND DISCOLORED DISPLAY SCREEN; THE DISPLAY SCREEN WAS HARD TO READ. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE DISPLAY SCREEN WAS DIM AND DIFFICULT TO READ IN THE LIGHT. THE REPORTER NOTED THAT THE DISPLAY SCREEN HAD SCRATCHES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165732 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |