FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3062675 · Received April 17, 2013

Report

Report Number
1416980-2013-09483
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
January 28, 2013
Report Date
March 28, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS CONFIRMED BY REVIEW OF THE LOGS. THE CAUSE OF THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW WAS DETERMINED TO BE INSUFFICIENT DRAIN, USE ERROR, TIDAL ULTRAFILTRATION (UF) REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. PG 12-30 HAS THE WARNING THAT "A TOTAL UF VOLUME SET TOO LOW CAN RESULT IN A GRADUAL BUILDUP OF UF VOLUME DURING THE THERAPY. THIS CAN RESULT IN AN IIPV SITUATION. " SECTION 13 "TOPIC 4: TIDAL THERAPY" GIVES THE TRAINER INSTRUCTIONS ON HOW TO PROPERLY SET TIDAL THERAPY VALUES. ON PG 13-9 THE INSTRUCTIONS STATE "A GOOD STARTING POINT WOULD BE, AT A MINIMUM, TO REVIEW A DRAIN HISTORY OVER THE PREVIOUS SEVEN TREATMENT DAYS. THE TOTAL AMOUNT OF UF OVER THE SEVEN DAYS SHOULD BE ADDED AND THEN DIVIDED BY SEVEN TO OBTAIN AN AVERAGE DAILY UF GOAL." THE SUGGESTED SETTING FOR TOTAL UF IS DESCRIBED ON PG 12-30 AS "SEVENTY PERCENT OF THE NORMAL NIGHT UF IS A GOOD STARTING POINT FOR DETERMINING THE OPTIMUM TOTAL UF. "

Description of Event or Problem · 1

DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING CYCLE 3. THE PATIENT'S LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 1900 ML AND THE DRAIN VOLUME WAS 3122 ML, WHICH MET IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166134 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 78 YR HOMECHOICE AUTOMATED PD SET WITH CASSETTE