FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3062672 · Received April 17, 2013

Report

Report Number
2531779-2013-04386
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
February 28, 2013
Report Date
March 20, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 05/30/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/06/2013 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS HEAVILY SCRATCHED AND AFTER THE PUMP POWERED ON, WAS FOUND TO BE DIM AND DISCOLORED. A TEST SCREEN WAS INSTALLED AND DISPLAYED PROPERLY.

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013 ALLEGING THAT THERE WAS A LINE THROUGH THE DISPLAY LENS. NO FURTHER INFORMATION IS AVAILABLE. THERE IS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS REPORT. THIS REPORT IS BEING REPORTED BECAUSE THE ALLEGED PRODUCT COMPLAINT REMAINED UNRESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166133 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1