FDA Adverse Event Malfunction Summary report: N

DIMENSION EXL WITH LM

MDR report key: 3062660 · Received April 17, 2013

Report

Report Number
1226181-2013-00176
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 22, 2013
Report Date
March 22, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K073604
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED (B)(4). DURING TROUBLESHOOTING, THE CUSTOMER INFORMED THE TSC SPECIALIST THAT THE SMOKE HAD COME FROM THE HEAT TORCH. THE TSC SPECIALIST ADVISED THE CUSTOMER TO REPLACE THE HEAT SOURCE, WHICH THE CUSTOMER DID. THERE WAS NO DAMAGE OBSERVED TO THE INSTRUMENT DUE TO THE HEAT TORCH OVERHEATING, AND NO SMOKE WAS OBSERVED AFTER THE HEAT TORCH WAS REPLACED. THE CAUSE OF THE SMOKE COMING FROM THE DIMENSION EXL WITH LM INSTRUMENT WAS A MALFUNCTION OF THE HEAT TORCH. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

THE OPERATOR OF A DIMENSION EXL WITH LM INSTRUMENT OBSERVED SMOKE COMING FROM THE DIAPHRAGM OF THE INSTRUMENT. THE OPERATOR RECEIVED A SLIGHT BURN WHEN HE UNLATCHED THE AREA. NO MEDICAL TREATMENT BEYOND FIRST AID WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166111 DIMENSION EXL WITH LM CLINICAL CHEMISTRY ANALYZER JJE SIEMENS HEALTHCARE DIAGNOSTICS DIMENSION EXL WITH LM