FDA Adverse Event Malfunction Summary report: N

POWEREASE¿ INSTRUMENTS

MDR report key: 3062616 · Received April 17, 2013

Report

Report Number
1030489-2013-01075
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 18, 2013
Report Date
April 24, 2013
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HBE
PMA / PMN Number
K111520
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO DAMAGE NOTED TO THE THREADS OF MAS HEAD THREADED INTERFACE. VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. DIMENSIONAL INSPECTION OF THE INNER SHAFT DIAMETER CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD, CONSISTENT WITH TORSIONAL OVERLOAD CONDITION. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.

Additional Manufacturer Narrative · 1

(B)(6). (B)(4). THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION; THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNSPECIFIED SPINAL PROCEDURE. IT WAS REPORTED THAT THE TIP OF THE SCREWDRIVER BROKE OFF WHILE RATCHETING THE SCREW FURTHER IN. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
165002 POWEREASE¿ INSTRUMENTS DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC SOFAMOR DANEK USA, INC NA CT11J066

Patients

Seq Age Sex Outcome Treatment
1