SITTER SELECT
Report
- Report Number
- 2020362-2013-00125
- Event Type
- Malfunction
- Date Received
- March 18, 2013
- Report Date
- February 26, 2013
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION OF THE RETURNED UNIT COULD NOT CONFIRM THE REPORTED ISSUE BUT FOUND BENT BATTERY SPRINGS. THE CONTACTS INSIDE THE NURSE CALL RECEPTACLE ARE DAMAGED, AS A RESULT THE CABLE WILL NOT GO IN COMPLETELY. DID NOT ATTEMPT TO CONNECT CABLE BY FORCE. SINCE NURSE CALL FUNCTION CANNOT BE UTILIZED, VOICE AND MUTE FUNCTIONS CANNOT BE TESTED. UNIT PASSES ALL OTHER FUNCTIONAL TESTS. NOTE: INSTRUCTIONS FOR USE STATE: HOLD ALARM VERTICALLY UPSIDE DOWN TO PREVENT BATTERY SPRING FROM BENDING DURING BATTERY INSTALLATION. INSERT FOUR NEW "AA" ALKALINE BATTERIES. TAKE CARE NOT TO DAMAGE BATTERY CONTACTS. TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. WHEN USING A NURSE CALL CABLE, ENSURE THE NURSE CALL CABLE IS PLUGGED IN TO BOTH THE ALARM AND THE WALL JACK BEFORE LEAVING THE PATIENT UNATTENDED. DO NOT STRETCH OR STRAIN CABLE TO AVOID POSSIBLE DAMAGE AND POSSIBLE MALFUNCTION. (B)(4).
CUSTOMER REPORTED THE ALARM SOUNDS CONTINUOUSLY WHEN USED WITH THE NURSE CALL CABLE AND THE PATIENT IS LYING STILL ON THE SENSOR. NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. NO PATIENT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 111595 | SITTER SELECT | KMI | J. T. POSEY CO. | 8361 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | NURSE CALL CABLE: ITEM #8282, LOT #UNK |