SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2013-00259
- Event Type
- Malfunction
- Date Received
- March 18, 2013
- Date of Event
- February 15, 2013
- Report Date
- February 18, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- 1314492-080712-002-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THE DEVICE EVALUATION CONFIRMED THE #2, #3, #4, #5, #6, #7, #8, #9 AND (.) KEYS TO BE INOPERABLE CAUSED BY A FAILED KEYPAD. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #2 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 12 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AD WILL BE DISPLAYED, INTERFERING WITH (B)(4) DRUG SEARCH). THE FAILED KEYPAD WAS REPLACED. BAXTER'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE #2 KEY ON A PUMP KEYPAD IS CONSTANTLY DISPLAYING. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 112752 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |