FDA Adverse Event Malfunction Summary report: N

OT VERIO TEST STRIPS

MDR report key: 3062371 · Received April 16, 2013

Report

Report Number
3008382007-2013-08055
Event Type
Malfunction
Date Received
April 16, 2013
Report Date
March 19, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K093745
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (6/24/2013) THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT PASSED TESTING. THE TEST STRIPS WERE EVALUATED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED; HOWEVER, A SECONDARY ISSUE WAS NOTED WHERE ERROR 4 WAS OBSERVED DURING CONTROL SOLUTION TESTS. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

THE LFS PRODUCT(S) HAS BEEN REQUESTED FOR RETURN TO LIFESCAN FOR EVALUATION. IF THE SUBJECT PRODUCT IS RETURNED, AN EVALUATION WILL BE COMPLETED AND THE FINDINGS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING "THE CLEAR PLASTIC PART NEAR THE PRONGS SEEMS TO COME DOWN A BIT FURTHER ON STRIPS." THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163471 OT VERIO TEST STRIPS GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3387609

Patients

Seq Age Sex Outcome Treatment
1 48 YR