OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-08042
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Report Date
- March 19, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (08/15/2013)-PRODUCT EVALUATION: THE ANALYSIS OF THE SUBJECT METER WAS COMPLETED ON 08/08/2013 BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE REPORTED COMPLAINT COULD NOT BE REPRODUCED. THE METER MET SPECIFICATIONS FOR TESTING. NO ISSUES WERE IDENTIFIED DURING INVESTIGATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE LFS PRODUCT(S) HAS BEEN REQUESTED FOR RETURN TO LIFESCAN FOR EVALUATION. IF THE SUBJECT PRODUCT IS RETURNED, AN EVALUATION WILL BE COMPLETED AND THE FINDINGS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN CANADA, ALLEGING INACCURATE RESULTS COMPARED TO THE SAME METER WITH READINGS THAT WERE "1 TO 2.5 MMOL/L" DIFFERENT. THE REPORTER DID NOT PROVIDE THE SPECIFIC READINGS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163469 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3386298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |