COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 3062311 · Received April 16, 2013

Report

Report Number
1061932-2013-00653
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
March 19, 2013
Report Date
March 19, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THE TUBING THROUGH PINCH VALVE PV37 WAS CUT. THE FSE REPLACED THE TUBING AND RESOLVED THE FLUID LEAK. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED LESS THAN TWENTY (20) MILLILITERS OF CLENZ CLEANING SOLUTION LEAKED OUTSIDE THE INSTRUMENT WITH LYSE TEMPERATURE SYSTEM ERRORS INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT GENERATED AS SAMPLES WERE NOT ANALYZED AT THE TIME OF THE EVENT. THE LABORATORY HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE INSTRUMENT WAS NOT IN USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163376 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1