ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-04349
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Report Date
- March 24, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 07/01/2013- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE LIFTED NEAR THE UP AND DOWN ARROW BUTTONS. DURING TESTING, ALL KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE DISCOLORED. A TEST SCREEN WAS INSERTED AND FUNCTIONED APPROPRIATELY.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE KEYPAD BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 162287 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |