FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3062300 · Received April 16, 2013

Report

Report Number
2531779-2013-04377
Event Type
Malfunction
Date Received
April 16, 2013
Report Date
March 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX HISTORY REVEALED POWER INTERRUPTIONS ON (B)(4) 2013. A ¿LOW BATTERY¿ AND ¿REPLACE BATTERY¿ ALARM WAS NOTED IN PUMP HISTORY. THE PUMP WAS SUSPENDED FOR 1 WEEK BETWEEN (B)(4) 2013. A VISUAL INSPECTION REVEALED NO DAMAGE OR MOISTURE TO THE PUMP. THERE WAS NO DAMAGE TO THE BATTERY CAP; ELECTRICAL CONNECTION WAS MAINTAINED WITH NO ISSUES NOTED. THE PUMP POWERED ON TO THE ¿VERIFY¿ SCREEN WITH NO ISSUES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES. A ¿LOW BATTERY¿ AND ¿REPLACE BATTERY¿ ALARM WAS INDUCED AND THE PUMP ALARMED APPROPRIATELY. THE REPORTED POWER ISSUE WAS NOT DUPLICATED DURING INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2013, THE DISTRIBUTER CONTACTED ANIMAS STATING THE PUMP'S DISPLAY SCREEN WAS BLANK. THE PUMP REPORTEDLY EMITTED AN "EMPTY BATTERY" ALARM. THE BATTERY WAS REPORTEDLY REPLACED AND THE PUMP POWERED ON WITH NO ISSUES. IT WAS NOTED THAT THE PATIENT'S BLOOD GLUCOSE MEASUREMENT WAS 4.254 G/L. THE REPORTED BG IS NOT INDICATIVE OF A SERIOUS INJURY AND DOES NOT MEET THE ANIMAS CRITERIA OF AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED DISPLAY SCREEN ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162286 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1