FDA Adverse Event
Malfunction
Summary report: N
EXTRAPURG PORTABLE REVERSE OSMOSIS MACHINE
MDR report key: 30623
·
Received February 29, 1996
Report
- Report Number
- MW1008488
- Event Type
- Malfunction
- Date Received
- February 29, 1996
- Report Date
- February 29, 1996
- Manufacturer
- BAXTER HEALTHCARE CORP.
- Product Code
- JRS
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THIS MACHINE IS SUPPOSED TO CLEAR CHLORINE AND CHLORAMINE FROM WATER BUT IT ONLY CLEARS CHLORINE. THE CARBON FILTER DOES NOT HAVE THE CAPACITY TO CLEAR THE CHLORAMINE. WHEN THE TECH NOTICED THE PROBLEM HE CALLED THE MFR AND THE CARBON FILTER WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EXTRAPURG PORTABLE REVERSE OSMOSIS MACHINE | WATER PURIFIER | JRS | BAXTER HEALTHCARE CORP. | DM-70 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |