FDA Adverse Event Malfunction Summary report: N

EXTRAPURG PORTABLE REVERSE OSMOSIS MACHINE

MDR report key: 30623 · Received February 29, 1996

Report

Report Number
MW1008488
Event Type
Malfunction
Date Received
February 29, 1996
Report Date
February 29, 1996
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
JRS
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THIS MACHINE IS SUPPOSED TO CLEAR CHLORINE AND CHLORAMINE FROM WATER BUT IT ONLY CLEARS CHLORINE. THE CARBON FILTER DOES NOT HAVE THE CAPACITY TO CLEAR THE CHLORAMINE. WHEN THE TECH NOTICED THE PROBLEM HE CALLED THE MFR AND THE CARBON FILTER WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTRAPURG PORTABLE REVERSE OSMOSIS MACHINE WATER PURIFIER JRS BAXTER HEALTHCARE CORP. DM-70

Patients

Seq Age Sex Outcome Treatment
1 *