FDA Adverse Event Injury Summary report: N

DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

MDR report key: 3062236 · Received April 16, 2013

Report

Report Number
2649622-2013-05078
Event Type
Injury
Date Received
April 16, 2013
Report Date
February 25, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. DEVICE: 5076 IMPLANTABLE PACING LEADS (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXPERIENCED TWAVE OVERSENSING WHEN PACED. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND REPLACED WITH A NEW PACE/SENSE LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162905 DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6944-58

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R D274DRG IMPLANTABLE DEFIBRILLATOR