FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3062071 · Received April 16, 2013

Report

Report Number
2032227-2013-01419
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
April 12, 2013
Report Date
April 12, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH PRIME ALARMS DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP GAVE AN ALARM DURING THE PRIME/REWIND PROCESS. TROUBLESHOOTING REVEALED THAT THE DRIVE SUPPORT CAP WAS PROTRUDING FROM THE CASE. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP. THE REPORTED BLOOD GLUCOSE READING WAS 6.8 MMOL/L. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163560 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LCAL

Patients

Seq Age Sex Outcome Treatment
1