Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 51 ML OF SOLUTION IN THE RESERVOIR. THE REPORTED CONDITION OF NON-DELIVERY WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE DEVICE SHOWED NO SIGNS OF BLOCKAGE THAT COULD HAVE CAUSED THE REPORTED CONDITION. FLOW WAS READILY OBSERVED AT THE LUER. FLOW CONTINUED WITHOUT STOPPING UNTIL THE SOLUTION WAS EMPTIED FROM THE RESERVOIR. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INFUSOR HAD DELIVERY STOP DURING PATIENT USE. THE DEVICE WAS FILLED WITH A SOLUTION OF FOLFOX MEDICATION. FORCE PRIMING WAS SUCCESSFUL BEFORE THE INFUSOR WAS CONNECTED TO THE PATIENT. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163939 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12H024

Patients

Seq Age Sex Outcome Treatment
1 FOLFOX MEDICATION