INFUSOR
Report
- Report Number
- 1416980-2013-09421
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Date of Event
- March 22, 2013
- Report Date
- March 26, 2013
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 51 ML OF SOLUTION IN THE RESERVOIR. THE REPORTED CONDITION OF NON-DELIVERY WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE DEVICE SHOWED NO SIGNS OF BLOCKAGE THAT COULD HAVE CAUSED THE REPORTED CONDITION. FLOW WAS READILY OBSERVED AT THE LUER. FLOW CONTINUED WITHOUT STOPPING UNTIL THE SOLUTION WAS EMPTIED FROM THE RESERVOIR. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
IT WAS REPORTED THAT AN INFUSOR HAD DELIVERY STOP DURING PATIENT USE. THE DEVICE WAS FILLED WITH A SOLUTION OF FOLFOX MEDICATION. FORCE PRIMING WAS SUCCESSFUL BEFORE THE INFUSOR WAS CONNECTED TO THE PATIENT. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163939 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 12H024 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FOLFOX MEDICATION |