LEAD MODEL 302
Report
- Report Number
- 1644487-2013-01034
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Date of Event
- March 19, 2013
- Report Date
- March 19, 2013
- Manufacturer
- CYBERONICS INC
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
MANUFACTURER REVIEWED X-RAYS OF IMPLANTED DEVICE. X-RAYS REVIEWED BY THE MANUFACTURER, NO GROSS LEAD DISCONTINUITIES VISUALIZED. DEVICE FAILURE OCCURRED BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
THE EXPLANTED LEAD AND GENERATOR WERE RETURNED ON (B)(4) 2013 WITH A RETURNED PRODUCT FORM STATING THAT THE DEVICE WAS EXPLANTED DUE TO LEAD DISCONTINUITY. ANALYSIS OF THE GENERATOR SHOWED THAT THE DEVICE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. ANALYSIS IN THE PA LAB CONCLUDED PROPER FUNCTIONALITY OF THE PULSE GENERATOR AND THAT NO ABNORMAL PERFORMANCE OR ANY OTHER TYPE OF ADVERSE CONDITION WAS FOUND. ANALYSIS OF THE LEAD SHOWED THAT THE LEAD DISCONTINUITY ALLEGATION WAS CONFIRMED. A SMALL PORTION OF THE LEAD ASSEMBLY (BODY) WAS NOT RETURNED FOR ANALYSIS AND THEREFORE A COMPLETE EVALUATION COULD NOT BE PERFORMED ON THE ENTIRE LEAD PRODUCT. DURING THE VISUAL ANALYSIS OF THE RETURNED PORTION THE POSITIVE WHITE ELECTRODE QUADFILAR COIL APPEARED TO BE BROKEN AT THE PROXIMAL END OF THE ANCHOR TETHER. SCANNING ELECTRON MICROSCOPY (SEM) WAS PERFORMED AND IDENTIFIED THE AREA AS BEING MECHANICALLY DAMAGED WITH PITTING WHICH PREVENTED IDENTIFICATION OF THE COIL FRACTURE TYPE. PITTING WAS OBSERVED ON THE COIL SURFACE. IT IS BELIEVED THAT STIMULATION WAS PRESENT FOR A CERTAIN PERIOD OF TIME AS EVIDENCED BY THE PRESENCE OF METAL PITTING. LOW MAGNIFICATION SEM ANALYSIS OF THE QUADFILAR COIL SHOWS CHARACTERISTICS TYPICAL OF A LEAD DISCONTINUITY. WITH THE EXCEPTION OF THE OBSERVED DISCONTINUITY, THE CONDITION OF THE RETURNED LEAD PORTIONS IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. NO OTHER OBVIOUS ANOMALIES WERE NOTED. THE RESISTANCE MEASUREMENTS TAKEN DURING DECONTAMINATION VERIFIED AN ELECTRICAL AND MECHANICAL CONTACT BETWEEN THE GENERATOR AND CONNECTOR PIN AT ONE POINT IN TIME. CONTINUITY CHECKS OF THE RETURNED LEAD PORTIONS WERE PERFORMED, DURING THE VISUAL ANALYSIS, AND NO OTHER DISCONTINUITIES WERE IDENTIFIED. X-RAYS WERE RECEIVED AND REVIEWED: THE GENERATOR IS VISUALIZED IN A NORMAL PLACEMENT IN THE LEFT UPPER CHEST. THE FILTER FEED-THROUGH WIRES APPEAR TO BE INTACT, AND THE LEAD CONNECTOR PINS APPEARS TO BE FULLY INSERTED INTO THE GENERATOR CONNECTOR BLOCK. THE LEAD IS VISIBLE, APART FROM A SECTION OF THE LEAD THAT IS BEHIND THE GENERATOR. THE LEAD WIRES APPEAR INTACT AT THE CONNECTOR PINS. A LEAD BREAK WAS NOT OBSERVED IN THE ASSESSED PORTION. REVIEW OF PROGRAMMING HISTORY SHOWED THAT A SYSTEM DIAGNOSTIC ON (B)(6) 2012 WAS WITHIN NORMAL LIMITS; HOWEVER, DIAGNOSTIC ON (B)(6) 2013 INDICATED HIGH IMPEDANCE. THE DEVICE WAS DISABLED ON (B)(6) 2013.
REPORTER INDICATED A PATIENT HAD HIGH LEAD IMPEDANCE WITH VNS DIAGNOSTICS TESTING ON (B)(6) 2013. THE VNS WAS DISABLED AND THE PATIENT WAS SENT FOR X-RAYS. SURGERY TO REVISE THE VNS APPEARS LIKELY, BUT HAS NOT OCCURRED TO DATE. ATTEMPTS FOR ADDITIONAL INFORMATION ARE IN PROGRESS.
AN IMPLANT CARD WAS RECEIVED TO THE MANUFACTURER INDICATING DEVICE DIAGNOSTICS WERE WITHIN NORMAL LIMITS WITH THE NEW VNS GENERATOR AND LEAD THAT WAS IMPLANTED ON (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163929 | LEAD MODEL 302 | LEAD | LYJ | CYBERONICS INC | 302-20 | 2239 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |