ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2013-00684
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Date of Event
- April 15, 2013
- Report Date
- May 31, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE, MOISTURE ENTERED THE INSULIN PUMP AND DESTROYED THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON IS PERMANENT ACTIVE DUE TO AN EXTERNAL MECHANIC DAMAGE.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
ON (B)(6) 2013, IT WAS REPORTED THAT THE PATIENT'S INFUSION DEVICE DISPLAYED E8 (POWER INTERRUPT) AND ALL OF THE BUTTONS ON THE DEVICE ARE WITHOUT FUNCTION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163926 | ACCU-CHEK SPIRIT | INSULIN INFUSION DEVICE | LZG | ROCHE DIABETES CARE AG | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |