FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT

MDR report key: 3061813 · Received April 16, 2013

Report

Report Number
2183996-2013-00679
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
December 13, 2012
Report Date
June 12, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATION REGARDING THE CUSTOMER'S ALLEGATION. THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT. THE ALARM FUNCTIONS OF THE PUMP WERE TESTED AND MEET THE SPECIFICATIONS.

Description of Event or Problem · 1

PATIENT REPORTED THE INSULIN DELIVERY OF THE INFUSION DEVICE IS INACCURATE AND THIS HAS RESULTED IN ERRATIC BLOOD GLUCOSE OF 2.0-22.0 MMOL/L (36-396 MG/DL) SINCE (B)(6) 2012. HE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE BUT REPORTED A NURSE ALSO BELIEVES THE DELIVERY IS INACCURATE. THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163528 ACCU-CHEK ® SPIRIT INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006862 NA

Patients

Seq Age Sex Outcome Treatment
1 079 YR