FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3061754 · Received March 29, 2013

Report

Report Number
3008642652-2013-00850
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 6, 2013
Report Date
March 26, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THE PINS IN THE ELECTRODE BELT TRUNK CABLE CONNECTOR WERE BENT. THE CAUSE FOR THE BENT PINS CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY DUE TO THE CONNECTOR BEING FORCED INTO THE MONITOR WHILE THE PINS WERE MISALIGNED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) WAS FOUND TO HAVE BENT PINS IN THE ELECTRODE BELT TRUNK CABLE CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129736 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA