FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3061711 · Received March 29, 2013

Report

Report Number
3008642652-2013-00842
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 1, 2013
Report Date
March 26, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE PINS ON BATTERY CHARGER POWER SUPPLY CONNECTOR WERE RECESSED, WHICH PREVENTED THE BATTERY CHARGER/MODEM FROM POWERING UP. THE ROOT CAUSE FOR THE RECESSED PINS COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, BATTERY CHARGER/MODEM SN (B)(4) WAS UNABLE TO POWER UP. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129730 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA