FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3061696 · Received March 29, 2013

Report

Report Number
3008642652-2013-00864
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 5, 2013
Report Date
March 27, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (MONITOR WON'T POWER UP) WAS CONFIRMED. UPON INVESTIGATION THE MONITOR WAS UNABLE TO BOOT UP PAST THE SPLASH SCREEN. THE CAUSE FOR THE POWER-UP FAILURE WAS ISOLATED TO A DEFECTIVE SD CARD. THE ROOT CAUSE FOR THE DEFECTIVE SD CARD COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE SD CARD. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULDN'T FULLY BOOT UP. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129580 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR