FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3061696
·
Received March 29, 2013
Report
- Report Number
- 3008642652-2013-00864
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 27, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (MONITOR WON'T POWER UP) WAS CONFIRMED. UPON INVESTIGATION THE MONITOR WAS UNABLE TO BOOT UP PAST THE SPLASH SCREEN. THE CAUSE FOR THE POWER-UP FAILURE WAS ISOLATED TO A DEFECTIVE SD CARD. THE ROOT CAUSE FOR THE DEFECTIVE SD CARD COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE SD CARD. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULDN'T FULLY BOOT UP. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129580 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |