LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00859
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- March 7, 2013
- Report Date
- March 27, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATIN, THE CHARGER/MODEM WOULD NOT POWER ON. THE CAUSE OF THE INABILITY TO POWER ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD SN (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT BGA CONNECTION, WHICH WAS DISCOVERED THROUGH THE USE OF LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THRUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN WAS SUBMITTED FOR FDA REVIEW ON 02/11/2013 AND IS CURRENTLY UNDER REVIEW. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CHARGER/MODEM. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM SN (B)(4) WOULD NOT POWER ON. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 131256 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |