FDA Adverse Event
Malfunction
Summary report: N
HX2 TEMPERATURE MANAGEMENT SYSTEM
MDR report key: 3061629
·
Received April 11, 2013
Report
- Report Number
- 1828100-2013-00393
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 20, 2013
- Report Date
- March 20, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEM CORP
- Product Code
- DWC
- PMA / PMN Number
- K071521
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) ISOLATED THE SOURCE OF THE HIGH LEAKAGE CURRENT TO THE CHANNEL A HEATER. THE FSR REPLACED THE HEATER AND THE UNIT OPERATED TO MANUFACTURER'S SPECIFICATIONS. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
Description of Event or Problem · 1
THE FIELD SERVICE REPRESENT (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, A HIGH REVERSE LEAKAGE CURRENT WAS NOTED. SINCE THE EVENT OCCURED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155709 | HX2 TEMPERATURE MANAGEMENT SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEM CORP | 809810 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |