FDA Adverse Event Malfunction Summary report: N

HX2 TEMPERATURE MANAGEMENT SYSTEM

MDR report key: 3061629 · Received April 11, 2013

Report

Report Number
1828100-2013-00393
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 20, 2013
Report Date
March 20, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEM CORP
Product Code
DWC
PMA / PMN Number
K071521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) ISOLATED THE SOURCE OF THE HIGH LEAKAGE CURRENT TO THE CHANNEL A HEATER. THE FSR REPLACED THE HEATER AND THE UNIT OPERATED TO MANUFACTURER'S SPECIFICATIONS. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENT (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, A HIGH REVERSE LEAKAGE CURRENT WAS NOTED. SINCE THE EVENT OCCURED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155709 HX2 TEMPERATURE MANAGEMENT SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEM CORP 809810

Patients

Seq Age Sex Outcome Treatment
1