FDA Adverse Event Injury Summary report: N

JOSTENT GRAFTMASTER

MDR report key: 3061517 · Received April 16, 2013

Report

Report Number
2024168-2013-02322
Event Type
Injury
Date Received
April 16, 2013
Date of Event
March 21, 2013
Report Date
March 22, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
H000001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THE LOT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A PERFORATION OF A SAPHENOUS VEIN GRAFT TO THE RIGHT CORONARY ARTERY. AN UNKNOWN DEVICE CAUSED A PERFORATION AND A NON-ABBOTT STENT WAS USED IN AN ATTEMPT TO SEAL THE PERFORATION. IT WAS DECIDED TO TAKE THE PATIENT TO THE OPERATING ROOM FOR OPEN HEART SURGERY; HOWEVER, WHILE WAITING FOR THE OPERATING ROOM, IT WAS DECIDED TO TRY TO SEAL THE PERFORATION WITH A 3.5 X 16 MM GRAFTMASTER. THE GRAFTMASTER COULD NOT CROSS THE LESION SO THE PATIENT WAS TAKEN TO SURGERY WITHOUT SEALING THE PERFORATION. TWELVE HOURS POST SURGERY, THE PATIENT EXPIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163658 JOSTENT GRAFTMASTER CORONARY STENT GRAFT MAF AV-TEMECULA-CT 656719

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention