FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3061493 · Received March 27, 2013

Report

Report Number
3008642652-2013-00746
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
February 18, 2013
Report Date
March 14, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. UPON INVESTIGATION THE PACK HAD A DAMAGED BATTERY CONNECTOR. THE ROOT CAUSE FOR THE DAMAGED CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE LAST PT TO USE THIS BATTERY PACK DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SVC DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, THE CONNECTOR ON BATTERY PACK SN (B)(4) WAS DAMAGED. THE LAST PT TO USE THIS BATTERY DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
127132 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA