FDA Adverse Event Malfunction Summary report: N

SECURE 3 MED/SURG BEDOBS 01/13

MDR report key: 3061491 · Received April 16, 2013

Report

Report Number
0001831750-2013-03350
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
March 24, 2013
Report Date
March 25, 2013
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP SUBMITTED AS FURTHER INVESTIGATION DETERMINED THAT, EVEN THOUGH THE BED WAS STUCK IN TREND, IT COULD STILL ATTAIN THE LOWEST, FLAT POSITION. BEING STUCK IN TREND WOULD BE A CAREGIVER ANNOYANCE, HOWEVER, IT IS NOT LIKELY TO HARM THE PATIENT AS CPR COULD STILL BE ADMINISTERED, IF REQUIRED. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED. THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO SERIOUS INJURY OR DEATH IF IT WAS TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE BED WAS STUCK IN TREND.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOOTEND WAS STUCK IN AN ELEVATED POSITION DUE TO A DAMAGED CPU BOARD AND STRIPPED COUPLER. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162070 SECURE 3 MED/SURG BEDOBS 01/13 BED, AC-POWERED ADJUSTABLE HOSPITAL FNL STRYKER MEDICAL-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment