FDA Adverse Event Malfunction Summary report: N

RESOLUTION CLIP CLIPPING DEVICE

MDR report key: 3061430 · Received April 16, 2013

Report

Report Number
3005099803-2013-02714
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
March 26, 2013
Report Date
March 26, 2013
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MND
PMA / PMN Number
K040148
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE CLIP ASSEMBLY WAS CONNECTED TO THE BUSHING. THE PRONGS COULD NOT BE OPENED BECAUSE THEY WERE LOCKED INTO THE CAPSULE. THE CLIP ASSEMBLY COULD BE DEPLOYED AND ONE CLICK WAS HEARD. BASED ON THE EVALUATION CONDUCTED AND THE DETAILS OF THE COMPLAINT, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS. A SEARCH OF THE COMPLAINT DATABASE CONFIRMED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED BATCH.

Additional Manufacturer Narrative · 1

PATIENT AGE/DATE OF BIRTH IS UNKNOWN AND UNAVAILABLE; HOWEVER THE PATIENT WAS REPORTED TO BE OVER 18 YEARS OLD. (B)(4) FOR THE REPORTED EVENT OF CLIP FAILED TO RELEASE FROM CATHETER. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS; HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USE DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. THE CLIP WAS UNABLE TO BE REOPENED AND HAD TO BE PULLED OFF THE TISSUE. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED AS BEING FINE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USE DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. THE CLIP WAS UNABLE TO BE REOPENED AND HAD TO BE PULLED OFF THE TISSUE. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED AS BEING FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162470 RESOLUTION CLIP CLIPPING DEVICE LIGATOR, ESOPHAGEAL MND BOSTON SCIENTIFIC - MARLBOROUGH M00522611 ML000425C3

Patients

Seq Age Sex Outcome Treatment
1