SPECTRUM INFUSION PUMP
MDR report key: 3061382
·
Received March 26, 2013
Report
- Report Number
- 1314492-2013-00293
- Event Type
- Malfunction
- Date Received
- March 26, 2013
- Date of Event
- February 26, 2013
- Report Date
- February 26, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- 1314492-080712-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED BY BAXTER. THE EVALUATION CONFIRMED THAT WHEN THE 3RD SOFT KEY IS SELECTED, THE OK KEY IS ENTERED CAUSED BY A FAILED KEYPAD. THE FAILED KEYPAD WAS REPLACED. BAXTER'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 3RD SOFT KEY DOES NOT FUNCTION CORRECTLY. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.