FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3061382 · Received March 26, 2013

Report

Report Number
1314492-2013-00293
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
February 26, 2013
Report Date
February 26, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
1314492-080712-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED BY BAXTER. THE EVALUATION CONFIRMED THAT WHEN THE 3RD SOFT KEY IS SELECTED, THE OK KEY IS ENTERED CAUSED BY A FAILED KEYPAD. THE FAILED KEYPAD WAS REPLACED. BAXTER'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 3RD SOFT KEY DOES NOT FUNCTION CORRECTLY. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123174 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1