FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2

MDR report key: 3061353 · Received April 10, 2013

Report

Report Number
2242352-2013-00311
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 12, 2013
Report Date
March 18, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD ON THE DEVICE. A VISUAL INSPECTION DETERMINED THAT THE MIDDLE OF THE COLD JAW HAD A VERY THIN CRACK. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, IT WAS OBSERVED THAT THE HEMOPRO 2 JAW WAS BROKEN UPON OPENING THE PACKAGE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148981 VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25068451

Patients

Seq Age Sex Outcome Treatment
1 NA