FDA Adverse Event Malfunction Summary report: N

SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION

MDR report key: 3061343 · Received April 16, 2013

Report

Report Number
2015691-2013-19824
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
March 20, 2013
Report Date
March 21, 2013
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
LDF
PMA / PMN Number
K813521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE BIPOLAR PACING CATHETER WITH ATTACHED MONOJECT 1.3 CC LIMITED VOLUME SYRINGE WAS RETURNED FOR EVALUATION. THE BALLOON INFLATED CLEAR AND CONCENTRIC WITH 1.3CC AIR AND THE BALLOON REMAINED INFLATED FOR 5 MINUTES WITHOUT LEAKAGE. CONTINUITY TESTING WAS PERFORMED ON THE DISTAL AND THE PROXIMAL ELECTRODES AND THERE WERE NO OPEN, INTERMITTENT OR SHORT CONDITIONS OBSERVED. NO VISIBLE DAMAGE TO THE CATHETER BODY, BALLOON LATEX, WINDINGS OR RETURNED SYRINGE WAS FOUND. VISUAL EXAMINATIONS WERE PERFORMED UNDER MICROSCOPE AT 20X MAGNIFICATION AND WITH THE UNAIDED EYES. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE COMPLETED. THE COMPLAINT COULD NOT BE CONFIRMED DURING THE EVALUATION. NO INDICATION OF A MANUFACTURING DEFECT WAS NOTED DURING THE ANALYSIS. IT COULD NOT BE DETERMINED IF ANY CLINICAL OR PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE DEVICE EVALUATION IS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT "IT WAS UNABLE TO PACE DURING USE." THERE WERE NO PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT "IT WAS UNABLE TO PACE DURING USE." THERE WERE NO PATIENT COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163753 SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION PACING CATHETER LDF EDWARDS LIFESCIENCES, PR PE075F5

Patients

Seq Age Sex Outcome Treatment
1