FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP BACKPACK

MDR report key: 3061314 · Received April 10, 2013

Report

Report Number
1419937-2013-00203
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 17, 2013
Report Date
April 4, 2013
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER CORD WAS SENT TO THE CUSTOMER. THE CUSTOMER CALLED MEDELA CUSTOMER SERVICE AND STATED THE TRANSFORMER HOUSING IS BROKEN AND THE WIRES ARE EXPOSED. THE POWER CORD WAS REPLACED AND REQUESTED THE DEFECTIVE POWER CORD TO BE RETURNED FOR EVAL. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVAL/INVESTIGATION. THEREFORE, NO CONCLUSION CAN BE MADE AS TO THE CAUSE OF THE EVENT. IN THE FOLLOW-UP WITH THE CUSTOMER, SHE STATED THERE WAS A BREACH AT THE TRANSFORMER HOUSING AND IT FELL APART. SHE HAD USED THE PUMP FOR 4 MONTHS, ABOUT 2 TO 3 TIMES A DAY. THE POWER CORD IS PLUGGED/UNPLUGGED ABOUT 3 TIMES A DAY. SHE INDICATED SHE DID NOT WITNESS ANY SPARKING, FIRE, OR FLAME, AND THAT NO INJURIES WERE SUSTAINED AS A RESULT OF THIS ISSUE. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MFR TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MFR TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE TRANSFORMER HOUSING IS BROKEN AND THE WIRES ARE EXPOSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148671 PUMP IN STYLE ADVANCED BREAST PUMP BACKPACK HGX MEDELA, INC. 9207010/57062 REV M 1212

Patients

Seq Age Sex Outcome Treatment
1