FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3061285
·
Received April 10, 2013
Report
- Report Number
- 8020893-2013-00835
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 15, 2013
- Report Date
- March 15, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK DISPLAY WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB AND BACKLIGHT INVERTER PCBS. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 150030 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |