PLMA DVC V11.51 1 N
Report
- Report Number
- 9615050-2013-00674
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- January 1, 2013
- Report Date
- March 15, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- OTHER
Narratives
INITIAL TESTING OF THE DEVICE WAS CONDUCTED AT THE USER FACILITY BY THE FIELD SERVICE ENGINEER ON (B)(4) 2013. THE POWER CORD WAS RECEIVED ON (B)(4) 2013. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT THE BLADES ON THE MALE END OF THE AC POWER CORD PLUG WERE MELTED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "MALFUNCTION." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THE BLADES ON THE MALE END OF THE AC POWER CORD PLUG WERE MELTED ON THE TOP SIDES. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 153639 | PLMA DVC V11.51 1 N | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE: LIST #12097, SN UNK |